Device The BCU #1 that was first delivered to Florence (and was before in Heidelberg for SW development). It is without ADC board, and was always affected by prob...
Device We had this issue on BCU #1 and also on BCU #2 Description Quando si configura le 2 schede HVC, qualcosa va storto e la scheda di sotto non si configura...
Description the following code results in only diode 1 set to full current diode.setDiodeState(1,True) diode.setDiodeState(2,True) diode.setDiodeState...
Device This issues affects BCU #2 and maybe other BCUs since it seems to be a firmware issue Occurrence When Notes 20130312 First observed. Repeatable,...
Occurence When Notes 20130325 first noted Description Sequence: a,E,o,c = Camera is cooled and working, and we get frames. We stop during for a coupl...
Description See attached email thread: MoCon pupilMotorBlue was not responding, but it recovers when commanded through IDL? Main.LorenzoBusoni 27 Mar 2013
Occurence When notes 20130820 Description We got an error with opcode 202 on BcuNumber3 when writing into DSP memory during frame grabbing. This could...
Occurence When notes 20121113 Clearly reproducible on BcuNumber3 on LgswSystemSx Description After a "fresh" reset of the BCU the initialization goes...
Welcome to the ARGOS Laser Guide Star twiki pages This twiki collects all the informations related to the 2 ARGOS laser guide star systems. It is written for the ...