8:00: we start opertations at horizon to fix the
CalUnitSystem DiodeSource problem. It seems related to the low temperature in the C-ring rack: JZ puts some tape on the unit to reduce the cooling ventilation. Now in operating condition the temperature is about 20°C.
9:45: Gilles goes up to swap fibers (the fiber splitter is not installed yet)
TN: 20150221_165046 bad
CalUnitSystem spots on
PatCams
Then we center the spots on the
PatCams and we reduce coma:
HEXA |
Pos |
X |
-1.52 |
Y |
-5.57 |
Z |
4.25 |
Rx |
-262 |
Ry |
155 |
10:30: we find a misconfiguration problem in the Lgsw network, we could not talk with ASM. Fixed: DX ASM IP is 192.168.11.12 (adsecdx)
We load subapertures for
TCoG algo. REC LGSW only to center the hexapod and get the reference positions: 20141202_125700
Diodes currents are set to 250, 150, 200mA to get a 4000ADU/subap fluxes.
12:00: ??? maybe PBL with arbitrator ???
We recenter again manually:
HEXA |
Pos |
X |
-2.52 |
Y |
-3.77 |
Z |
4.3 |
Rx |
-470 |
Ry |
55 |
We try to close just on LGSW. We have the new Hexa coordinates:
HEXA |
Pos |
X |
-2.52 |
Y |
-3.77 |
Z |
4.35 |
Rx |
-471 |
Ry |
56 |
TN |
Notes |
Plot |
20150221_190749 |
NO disturb, LGSw only, G=0.1,0.1,0.1 |
|
20150221_190913 |
G=0.5,0.3,0.2 |
20150221_191120 |
no loop, to check Snapshot entries |
20150221_191549 |
disturb=1.2", G=0.6,0.5,0.4 |
20150221_191718 |
disturb=1.2", G=0.5,0.3,0.2 |
20150221_192500 |
disturb=1.2", G=0.7,0.7,0.6 |
We pause the loop and we center the FLAO board on the on-axis source.
We load the Combined REC (FLAO BIN2 + LGSW)
TCoG and we close the loop:
We are having PBL with the
CalWhiteLightSource: it disconnects and resets to closed shutter + power off positions...
TN |
Notes |
Plot |
20150221_200425 |
Disturb=1.2", G=2.0,0.5,0.4, no TS |
|
20150221_210945 |
Disturb=1.2", G=2.0,0.5,0.4, no TS |
|
20150221_211019 |
Disturb=1.2", G=2.0,0.5,0.4, no TS |
|
20150221_211035 |
Disturb=1.2", G=2.0,0.5,0.4, no TS |
|
20150221_211122 |
OL |
|
20150221_211135 |
OL |
|
We spend the afternoon in trying to debug the Truth Sensing software.
Flow meter installation
We installed a flowmeter on the
LgswUnitDx cable cover. It allows to tune the flow from the filter installed on the C-ring to the LGSW itself. We regulated it to get a 2l/min flux on the LGSW.
Calibration light sources
The DX green off-axis LED source was changed in the morning to a spare one (#6 -> #3). It could not be remotely operated anymore. However, the new one showed similar behavior. It was concluded that it is probably temperature. The lower part of the unit was covered with tape to avoid the cold air flowing directly through from the van below.
As mentioned above the white lite source showed some disconnects, which triggered to close the shutter after reconnection and also some flickering of the source, which looks like unstable power supply. Has to be monitored further!
Laser system test
Laser system was switched on and power was measured short after.
SX 16.82 17.31 15.47
DX 17.6 16.6 15.46
With the Arbitrator test in the late afternoon/early evening the laser have been switch on and of again.
DX laser 3 (Head 5) didn't open the internal shutter. After manual closing and opening with Ranger the laser was working properly.
LUCI support
LUCI motors have not been available through software. JZ supported the efforts to resolve the problem. A power supply was broken. After replacing with the spare there have still been some issue and not everything working. First switch cables but maybe still more. Open Issue.
--
MarcoBonaglia - 21 Feb 2015