9:00: we go up to crane back the LGSW: ML leads the operation.
12:00: installation completed, we go back to control room and we switch on the
LgswUnitDx.
New laser acquisition target
We measure the new
FoV centers on Patrol cameras:
WFS |
X |
Y |
Blue |
179 |
146 |
Yellow |
176 |
117 |
Red |
177 |
127 |
- Laser acquisition target tag: 20161015_192400
We also control and re-optimize the Patrol cameras clocking...
TNs |
Plot |
*DX: ['20161015_194932', ...] |
|
We try to close the loop using the May reconstructor, LGSW only. We flatten the shell using flat 20160515_ARGOS_10modes
Best pupil position is with motors in:
DX HEXA |
-5.16 |
-8.27 |
1.18 |
292 |
137 |
With IS we see the full pupils while with the cal source the bottom left part of yellow is vignetted (or fainter!) (left in figure below)
We compare the flux distribution with the one of May and we notice the same effect but on the Red at that time (right in figure below)
With a mag=10 we test again the Truth Sensing closing the AO loop with the Pyramid and the LGSW. We move the Pyramid Board to null the differential focus between the Pyramid WFS and the LGSW.
TN |
AO Gain |
AO CL |
TT Res. WF [nm rms] |
HO Res. WF [nm rms] |
AO rec |
DIMM ["] |
JCL |
PCL |
Notes |
20161015_173638 |
0.5,0.5,0.4 |
T |
123.496 |
56.393 |
20161013_160000 |
-1 |
T |
F |
no dist |
20161015_173704 |
0.5,0.5,0.4 |
T |
104.349 |
48.4 |
20161013_160000 |
-1 |
T |
F |
20161015_174323 |
0.6,0.5,0.4 |
T |
51.8505 |
60.231 |
20161013_160000 |
-1 |
T |
F |
1" dist |
20161015_174341 |
0.6,0.5,0.4 |
T |
63.5841 |
65.0478 |
20161013_160000 |
-1 |
T |
F |
20161015_174400 |
0.6,0.5,0.4 |
T |
52.9648 |
61.8985 |
20161013_160000 |
-1 |
T |
F |
20161015_174500 |
0.6,0.5,0.4 |
T |
65.9369 |
67.7154 |
20161013_160000 |
-1 |
T |
F |
APD calibrations
11:30 We connect the APD fibers to the AGW.
Snapshot with APD in the dark: 20161015_193107
We start the dark calibration. We perfom a set of 4 seconds measures.
APD1 |
APD2 |
APD3 |
APD4 |
138 |
213 |
3106 |
275 |
Too much light on APD3, we go up and check the fibers.
APD1 |
APD2 |
APD3 |
APD4 |
140 |
300 |
400 |
300 |
It seems ok but then we try to measure the flat and we measured:
APD1 |
APD2 |
APD3 |
APD4 |
8000 |
12000 |
1000 |
11000 |
We change the APD3 fiber (FIBER2) with the SPARE2 and we measured a flat of:
APD1 |
APD2 |
APD3 |
APD4 |
10000 |
15000 |
9000 |
13000 |
We measured the dark and the flat with 40 seconds measures:
|
APD1 |
APD2 |
APD3 |
APD4 |
dark [ph/s] |
142 |
280 |
820 |
348 |
flat [%] |
76 |
99 |
60 |
100 |
With the new dark and flat we close the AO loop with APD and LGSW. We use the Pyramid for the Truth Sensing with low flux (mag>18). We move the Pyramid Board to null the differential focus between the Pyramid WFS and the LGSW.
TN |
AO Gain |
AO CL |
TT Res. WF [nm rms] |
HO Res. WF [nm rms] |
AO rec |
DIMM ["] |
JCL |
PCL |
Notes |
20161016_000127 |
0.1,0.4,0.3 |
T |
1107.25 |
81.4256 |
20161014_235200 |
-1 |
T |
F |
|
20161016_000148 |
0.1,0.4,0.3 |
T |
1125.85 |
84.3245 |
20161014_235200 |
-1 |
T |
F |
|
20161016_000204 |
0.1,0.4,0.3 |
T |
1163.61 |
97.1494 |
20161014_235200 |
-1 |
T |
F |
|
20161015_230338 |
0.5,0.5,0.4 |
T |
2505.32 |
241 |
20161014_235200 |
-1 |
T |
F |
|
20161015_231356 |
0.1,0.4,0.3 |
T |
1202.54 |
33.9586 |
20161014_235200 |
-1 |
T |
F |
|
20161015_232527 |
0.1,0.4,0.3 |
T |
1388.8 |
34.0895 |
20161014_235200 |
-1 |
T |
F |
|
20161015_232559 |
0.1,0.4,0.3 |
T |
1308.09 |
34.0503 |
20161014_235200 |
-1 |
T |
F |
|
20161015_234739 |
0.1,0.4,0.3 |
T |
2191.83 |
42.8717 |
20161014_235200 |
-1 |
T |
F |
|
20161015_234805 |
0.1,0.4,0.3 |
T |
2192.53 |
43.2856 |
20161014_235200 |
-1 |
T |
F |
|
20161015_235218 |
0.1,0.4,0.3 |
T |
2179.27 |
43.2756 |
20161014_235200 |
-1 |
T |
F |
|
20161015_235432 |
0.1,0.4,0.3 |
T |
2692.42 |
47.34 |
20161014_235200 |
-1 |
T |
F |
|
20161015_235736 |
0.1,0.4,0.3 |
T |
1129.81 |
67.1931 |
20161014_235200 |
-1 |
T |
F |
|
20161015_235757 |
0.1,0.4,0.3 |
T |
1094.64 |
66.5177 |
20161014_235200 |
-1 |
T |
F |
|
The TS was able to converge to the same modal offset measured in the morning with a bright ngs.
ASM skip frame
After a few minutes of AO closed loop with the 1" disturbance we notice the Skip Frame flag and we check the temperature in the Housekeeper GUI of the ASM.