The three DX cells are refurbished and sitting now in the lab, waiting for packing. All three have been checked on transmission in a collimated 7.6mm beam and a converging/diverging beam using f=100mm lenses in the same beam. Applying no voltage pulses the suppression of the cells is around ~2000..3000 in a collimated beam. due to some ringing present this performance drops with our usual 2µs pulses at 10kHz to ~500. The average over all delays is hard to estimate in this setup. According data can be found here for reference: https://bscw2.mpe.mpg.de/bscw/bscw.cgi/d870828/15oct_2014_refurbished_cells.rar The attached table gives the average count values in the 7.6mm circle, as read from the screen.Performance_summary_refurbished_cells.xlsx: Performance_summary_refurbished_cells.xlsx
I | Attachment | Action | Size | Date | Who | Comment |
---|---|---|---|---|---|---|
![]() |
P1000256.JPG | manage | 3 MB | 04 Nov 2014 - 03:12 | MarcoBonaglia | |
![]() |
Performance_summary_refurbished_cells.xlsx | manage | 10 K | 04 Nov 2014 - 03:12 | MarcoBonaglia |