Occurence

When Notes
20130325 first noted

Description

Sequence: a,E,o,c => Camera is cooled and working, and we get frames. We stop during for a couple of hours during a videoconf (S,C) but camera is kept cold

We restart (c,s). We don't get any frame. 'L' shows problems. The log of wfsServer shows problems.

See the full email thread: Gmail_-_Camera_doesnt_restart_after_a_pause.pdf

The voltages applied to the camera are wrong: Screenshot-7.png

See also log_inaf_pnccd_pc_service.txt: log of wfsService from 20 to 25 March

Actions

When wfsService is restarted on the pnccd_pc the camera is back in operation. See the email: Gmail_-_Camera_doesnt_restart_after_a_pause.pdf

Try to reproduce it asap.

Status

The memory leak that caused the server to crash was identified and fixed in a couple of weeks.

BugsInfoForm edit

Title WFSCamera service memory leak
DevicesInvolved WfsCamera1
Tag
FirstSeenOn 25 Mar 2013
Status Closed
I Attachment Action Size Date Who Comment
Gmail_-_Camera_doesnt_restart_after_a_pause.pdfpdf Gmail_-_Camera_doesnt_restart_after_a_pause.pdf manage 149 K 26 Mar 2013 - 12:10 LorenzoBusoni email thread about camera sequencer issue
Screenshot-7.pngpng Screenshot-7.png manage 93 K 26 Mar 2013 - 12:10 LorenzoBusoni screenshot of wfsCameraController showing wrong voltages applied to the camera
log_inaf_pnccd_pc_service.txttxt log_inaf_pnccd_pc_service.txt manage 1 MB 26 Mar 2013 - 12:09 LorenzoBusoni log of wfsService from 20 to 25 March
This topic: ARGOSPublic > WebHome > BugsSummary > ThingsThatIDontUnderstand0006
Topic revision: 16 May 2013, LorenzoBusoni
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding AOWiki? Send feedback