Diary
20150820_23xxxx |
PockelsCell145 installed in BlueChannel replacing PockelsCell144 |
20150821_150055 |
PCT measurement |
20150821_163526 |
PCD off |
20150821_174238 |
Low flux regime, loops on. Snapshot |
20150821_174600 |
Chiller halted. Automatic poweroff |
20150821_18xxxx |
Test of MoCon ethernet port |
20150821_184600 |
Rack powercycle. Power up |
20150821_192318 |
Red pupil ear. Snapshot |
20150821_192429 |
PCD on |
20150821_192547 |
PCT measurement |
20150821_195819 |
Pupil stabilization loop test |
20150821_215603 |
PCD off |
20150821_224723 |
PCD on |
20150821_224924 |
Pupil stabilization loop test |
20150822_021200 |
Device Error Pupil Motor Red |
20150822_022500 |
Patrol Camera disconnected and re-plugged after about 30min. After that device error on MoCon: SW restarted |
20150822_170222 |
PCT measurement - Red Pupil Ear disappeared |
20150822_202506 |
PCT measurement. Ok |
20150822_202800 |
shutdown-restart-powerup |
20150822_205600 |
SW restart |
20150823_165520 |
PCT measurement. Ok |
20150823_170000 |
Poweroff. We need the electronics for dx-lgsw |
Zabbix
Zabbix patched to allow for 5min zoom in graphs. Downloaded a patch called zabbix-2.0.7-zoom.path from the web and installed on both
WorkStation1 and
WorkStation2
Pupil stabilization loop stability with flux variations
Between 20150821_195819 and 224924 we run with pupil and jitter loop closed and different values of flux to see if the algorithms have a dependence on the flux. The short answer is no.
The flux is adjusted, we wait for the loops to converge and we store pupil motor position and piezo mirror offset. We repeat for several flux levels. We never saw an abrupt change of pupil motor position or piezo mirror offset after the flux was adjusted, and this is considered an argument to assert that the pupil stabilization algorithm is robust to flux variations.
We could instead notice a slow drift with 1h timescale. The
PtV drift over a 2h range are the following:
Ch (x,y) |
Pupil Motor (steps) |
Piezo (mrad) |
Blue |
5, 7 |
0.2, 0.2 |
Yellow |
11, 12 |
0.5, 0.3 |
Red |
18, 17 |
0.5, 0.7 |
Pockels cells
System has been long all night long without flaws.
Pockels cells are still ok: they have been on continuosly from 20150821_000519 to 20150821_163526 and no issues in the PCT have been seen. We tested both normal humidity (30%) and low humidity regimes.
At 20150821_163526 we powered off the pockel cell driver to check the pupil stabilization routine in low flux regime
At 192000 after a complete power-cycle of the system the red pupil has "ears"
The system is left on continously most of the time with PCD on and dry air. At 20150822_170222 the "ears" are disappeared and don't reappear until the end of the test on 20150823_170000