How To Setup ARGOS for AO calibrations
Pre-requisits:
Organize telescope access with Telescope Manager and mail to telescopework
Telescope must be at Zenith - you can check it in the IIFGUI.
Stow pins must be in - you can check it in the MCSGUI.
HBS should be on - you can check it in the MCSGUI.
Authorize as LUCI, bentGregorianArgos (
HowToAuthorizeLUCIAtARGOS
) - you can check in the IIFGUI)
Set the Adsec Shell in Adsec Arbitrator GUI
Enable Retroreflector mode from the adsec terminal (enable_rr_mode.sh) - you can check in the AOSGUI
Upper floods must be switched off, lowers can be on
Now done with authorization as bentGregorianArgos:
Fiber switch must connect ARGOS
BCU
to Adsec - (
HowToUseTheFiberSwitch
) - (now done with authorization of bentGregorianArgos)
not yet fully:
Go
AdSec
Control GUI: Tab 'Focal station' -> Press 'Set'
Enable Argos Arbitrator in AOSGUI - (now done with authorization of bentGregorianArgos)
Procedure
Startup ARGOS and FLAO software, see
HowToStartSoftware
and
HowToStartPyramidWfs
Open Arbitrator GUI.
Put
LAS
and
LAN
in
simulation
Execute the sequence up to state "READY_TO_CALIBRATE_ON_CALSOURCE",
select therefore the proper Light source
!!
[Parallel] Deploy calibration unit, see
HowToSetupCalibrationUnit
[Parallel] Setup Hexapod: see
HowToEnableHexapodOffload
[Parallel] Switch on White light source (shutter position 100)
[Parallel] Switch on Diode sources (100mA current)
After finishing all previous tasks you should see the
CalSourceDiodes
on the Patrol cameras (Arbitrator GUI, tab "LGS Acquisition").
If not visible you have to search for them with the Hexapod
If you see large coma you can remove it following
HowToRemoveComa
.
If they look good you put them on the wavefront sensor
Open 'CCD viewer': Applications -> Argos ?x -> ccd viewer
Open Arbitrator GUI, tab "LGS Acquisition"
Select "hexapod & patrol camera"
Click in each of the camera displays on the corresponding spot of the diod
Press button "Acquire manually
Repeat until pattern on "CCD viewer" looks good
Update in PSFGUI the Hexapod "Global Offsets" with the values from "Platform Position"!