DT: training of XY on SW and HW
20:30: we deploy the cal unit and we try to do SX calibrations
SX HEXA |
2.79 |
0.32 |
-0.78 |
208 |
-308 |
New
modal histories for KL_v18, using modal amplitude: 20151211_164800_rms0.10.fits
We first try to acquire an IM using the old MH used for REC: 20151212_220600 and LGSW IM: 20151212_202229
We acquire a new flat: 20160517_033800_ARGOS_150modes
Modal amplitude: 20151211_162400_rms0.05, KL_v18, MH: 20160517_KL_v18_150_0.05
LGSW IM: 20160518_055735 it is not good, the lower border of the 3 pupils is bended...
LGSW REC: 20160518_055800, COMBO REC: 20160518_055801
Then we generate a new MH using KL_v20 and we acquire a new LGSW IM: 20160518_060514, LGSW REC: 20160518_062100, COMBO REC: 20160518_062101
We try again an acquisition using the new LGSW REC for KL_v20:
LGSW IM: 20160518_062548, LGSW REC: 20160518_064037, these have been done using a modal amplitude of 0.10
We try to increase the preamble from 100 to 500, nothing changes
We try to increase the light level and the number of cycles but nothing changes...
Then we try to increase the number of up and down steps from 3 to 10 (we are using KL_v18):
LGSW IM: 20160518_065724, LGSW REC: 20160518_065800, COMBO REC: 20160518_065801
And then just 6 steps:
LGSW IM: 20160518_070455
Then we try KL_v20:
LGSW IM: 20160518_071032, LGSW REC: 20160518_071100, COMBO REC: 20160518_071101
BCU stucks at 8:22:49 UTC, it continues to read duplicated ASM records
10:27:24 ring buffer exception timeout of 2.4s when reading 1 record or 6.0s when reading 10 records...