When | Where | Who | What |
---|---|---|---|
20130506 | OAA | JZ, GOX | Delivered |
20131011 | OAA | GOX | Upgrade of the electronic panel on the external side |
20140114 | OAA | MB, JZ, TM | Removed from LgswUnitSx. Installed on LgswUnitDx |
20140318 | LBT | GOX | Pumping camera (>2hrs) installed on the WFS DX table |
20140508 | LBT | GOX/LBTO | Installation of pressure regulator incl filter, right after the suppy valve |
20140509 | LBT | GOX | Flush with N2, installation of permanent pressure gauge, pumping (~30min) |
20150628 | LBT | GOX | camera pump out (~20min) |
20180210 | LBT | GR | Pressure was 9.3 mbar. Pumped down to 0.6 mbar. Last reading was 0.76 mbar |
When | What |
---|---|
20130305 | description of issue |
When | Where | Pressure from pressure gauge | Pressure camera sensor (Voltage reading, converted in mbar) |
Humidity raw value (water cooling only, at -20C, max value when warming up) |
Comments |
---|---|---|---|---|---|
20140509 | LBT | <1mbar | (0.088V,<0.02mbar ) | 0.9 | camera pumped out |
20141104 | LBT | 15mbar | (0.1V,1.2mbar ) @-20C | (1.14,1.05,1.249) | This is fully acceptable since it does not trigger any warning or error |
20150628 | LBT | 31mbar | - | - | maintenance, camera pump down to ~1mbar |
20151017 | LBT | 8.5mbar | - | - | - |
20150307 | LBT | 13mbar | ~2.4mbar @ 10C | - | - |
20161021 | LBT | 1 mbar | Pressure reading by TM | ||
20180210 | LBT | 9.6 mbar | - | - | Maintenance, camera pump down to 0.6 mbar |
![]() | |