LgswDichroicSystem check In the last days we found many issues with the movement of both the dichroic and fold mirror on the LgswDichroicSystemSX while on the Lgs...
If the spots are very broad you could try to limit the jitter loop stroke to avoid them to jump on the next subaperture. Anyhow before you probably would try to o...
19:00 22:00: time allocated to LUCI/AO G111.25 * Problem with the sign in the stages directions signs, solved * Pointing reference position for FLAO is 7...
LgswUnitDx work 8:30: WG and MR go to check the cal unit illumination patter. It seems ok, maybe an alignment problem? 10:20: BcuNumber3 fails to connect. We must...
Description The Tip Tilt System is composed by: * An ApertureWheel used to set the Tip Tilt system FoV * A QuadCellUnit installed on the technical viewer ar...
The documentation about how to start/stop, restart, and properly power off the LBT AdSec is maintained in the LBTO Twiki at the current page: http://wiki.lbto.org...
Description This is the new model of PockelsCells rebuild after the OzoneProblem, upgraded to version 2.1 (see PockelsCells) Known history When Where Who ...
Visual Inspection 11:00: telescope back to Zenith, we climb and check the racks: fine. The pressure gauge on SX PNCCD had low battery. We pumped it anyway up to 0...
Visual check 15:00: we climbed on telescope and had a visual check of the LGSWs and LAS boxes status. * The 2 LGSWs appear very muddy, a lot of mud water accum...
Orientation on the DX side See email to Gilles and Lorenzo of 28/02/2015: So from our evaluation the LUCI CCD top is vertical in the telescope system when the rot...
Main.SebastianRabien 12 Aug 2016 Julian and SR on the mountain. Actions performed: * The new remote power is installed and tested. New engineering gui installe...
Currently, the workstation is located in Heidelberg (Faraday) and used as backup system. Identification * Dell R520 * On the front side there is a plastic ...
Currently, the workstation is located in Heidelberg (Faraday) and used for testing the software in simulation. Identification * Dell R520 * On the front si...
This is a spare server currently located in Heidelberg (Faraday). Currently, the workstation is used for testing the software in simulation. Identification DELL ...
i. Open LGSW Lab GUI i. In the Dichroic tab select the Expert tab i. In the Command white field write the following command: Command Istruction 00PS...
Identification Dell R510 Setup as dx lalas. On the rear side close to the 2 power supplies there is a label with the service tag: ST38XJB5B and the Express Servic...
Identification Dell R510 Setup as sx lalas On the rear side close to the 2 power supplies there is a label with the service tag: ST48XJB5J and Express Serice Tag:...
Identification DELL R520 Setup as SX LGSW On the rear side close to the 2 power supplies there is a label H6W5ZX1. This is the service tag. The Express Service Ta...
Identification Dell R520 Setup as DX LGSW. On the rear side close to the 2 power supplies there is a label CSC2H5J (27834711895). This is the service tag. On the ...
03:00UT: Si sale a controllare il DichroicSX che non si muove correttamente. Non si capisce bene, sembra che il Fold Mirror duri fatica a partire. Nel pomeriggio ...
Fast guiding check w/ oscilloscope 11:30: Tommaso and Sebastian go on the telescope to check the Fast Guiding delay. We connect an oscilloscope after the 10x divi...
TT w/ Pyr Bin4 test 8:00: we start with PyrWfs, setup as Bin4, mag 15. * CCD47 is not set to Bin16 during the preset * TS at Bin4 is missing the KL basis at...
Main.SebastianRabien 15 Dec 2015 When to exchange a laser head? A laser heads needs to be exchanged if it is broken. This is the case if a laser shows very low p...
Late opening for technical reasons SDSSJ1110 6459 04:10UT Loop closed 04:24UT We keep on being visited by airplanes 04:45UT Un po di snapshot al passare delle n...
LgswUnitSx and LgswUnitDx HW servicing 8:00: we unpack the PCD and the amplifier from 3L boxes, we bring it to 5. We also spend 1:30h in searching for the WfsCame...
This is the pump that is needed to vacuum the WfsCamera Checking camera housing pressure and pumping it out Reading the pressure * (the software pressure meas...
BCU SX debugging 15:00: we start to check the behaviour of the SX BCU. We notice that the main issue is when the Pyramid slopes are enabled. On DX we measure slop...
The BCU is acting as: frame grabber, slope computer, High Voltage Controller for the tip tilt mirrors, collector of slopes from the FLAO BCU and from the TipTiltU...
This procedure is used to match the 3 LGS images produced by the CalUnitSystem to the LgswUnit entrance FoVs. It is composed by a measurement procedure that scans...
5:30: Gustavo deploys the arm, we have pbl starting AdSecSx: it periodically faild while loading the DSP program. We try many times to power cicle. LB finally can...
LgswUnitDx FoV adjustment to CalUnit We adjusted the position of the LGSW EW to match the cal unit sources, below a summary of the measurements: * 20160312_DX ...
13:30: Arrival at telescope 16:00: we try to start up both systems. Due to power outage the SX LGSW computer was off, we turned it on again. The DX LGSW computer ...